Volume 23, Issue 6 , Pages 753-767, December 2009
Medications that distort in vitro tests of thyroid function, with particular reference to estimates of serum free thyroxine
The combination of serum thyroid-stimulating hormone (TSH) with measurement of circulating thyroid hormones greatly improves sensitivity and specificity of thyroid diagnosis, but these assays are not impeccable. Estimation of serum free T4 conveniently accommodates variations in the concentration of thyroxine-binding globulin (TBG), but no current technique reliably reflects the in vivo free T4 concentration in numerous other situations. The effect of circulating competitors that increase T4 and T3 in vivo, in particular, many medications, is under-estimated by current free hormone estimates that involve sample dilution. Non-esterified fatty acids generated during sample storage and incubation can spuriously increase the measured free T4 estimate, especially after in vivo treatment with heparin. These artefacts are unlikely to be overcome by current assay strategies. Total serum T4, corrected for alterations in TBG concentration, gives a more robust estimate of thyroxine concentration than current methods of free hormone estimation and should now be reintroduced as the ‘gold standard’.
Keywords: free T4 interference, sample dilution, aspirin, furosemide, carbamazepine, non-steroidal anti-inflammatory drugs, non-esterified fatty acids, heparin, serum total T4
To access this article, please choose from the options below
PII: S1521-690X(09)00073-6
doi:10.1016/j.beem.2009.06.004
© 2009 Elsevier Ltd. All rights reserved.
Volume 23, Issue 6 , Pages 753-767, December 2009